Iontophoresis Device Idromed4 - Preparation of the treatment

Impressum/Imprint      Sweat side DeutschSweating site francais

- Index - Order form - Iontophoresis forum -   Manual - how to use iontophoresis -

Extreme Sweating is not normal. Hyperhidrosis or Excessive Sweating treatment!

This is a sample of the content of the manual from the iontophoresis-device Idromed4 - Note: we cannot guarantee the actuality and accuracy.

If you buy a Idromed iontophoresis device a actual manual is included.

The manual is spreaded over several pages.

With the idromed® 4 you have purchased a tried and tested iontophoresis unit for the treatment of excessive sweating (hyperhidrosis) on hands, feet and armpits (with optional accessories). We hope that you will be satisfied with the use of the idromed® 4. In order to achieve good results, please follow these instructions exactly.

CONTENT

1.    Important information on the idromed® 4   
1.1.    Safety information 
2.    Scope of delivery   
3.    Application of the idromed® 4  
3.1.    Fields of application  
3.2.    Preparation of the treatment   
3.3.    Treatment
3.3.1.    Treatment of hyperhidrosis on the hands  
3.3.2.    Treatment of hyperhidrosis on the feet   
3.3.3.    Treatment of hyperhidrosis on hands and feet   
3.3.4.    Treatment of hyperhidrosis at the armpits   
4.    Important tips and information  
5.    Cleaning and Care   
6.    Maintenance   
7.    Problems and how to solve them   
8.    Technical Data  
9.    Warranty and liability  

3.2. Preparation of the treatment

The idromed® 4 is used with a power pack. After the following preparations your unit is ready for use:
a) For the treatment you should separate the two parts of the suitcase. To do this, open the suitcase completely, put it on a plain surface and separate the 2 parts carefully in opposite directions (see arrows on the hinges on the suitcase).
b) Fill the 2 treatment tubs with tap water to a depth of approx. 3-4 cm. (the finger and toe joints should be just covered).
c) Connect one of the cables to one of the plate electrodes and to the (+)-connection on the front of the idromed® 4.
d) Connect the 2nd cable of the electrode with the other electrode and with the (-)-connection on the front of the idromed® 4.
e) Place the electrodes 1 in each tub and cover the electrodes with the distance grids.
f) care, that the black control knob for the regulation of the current intensity is pointing to [1] and that the foot switch is pointing to [0].
g) Connect the plug of the power supply to the rear of the idromed® 4 and plug the power supply to a mains socket (220-230 V / 50 Hz). The green control lamp on the power pack is now alight.
The idromed® 4 is now ready for the treatment!
The LCD-display of the idromed® 4 indicates a positive value only when the current flow is closed via the body. Only then the really flowing therapy current is indicated on the display.
When the 2 electrodes have no contact to the body and the foot switch is switched to 1, the LCD-display always indicates the value [0]! (see also Chapter 7)